Product Overview
- DIODE LASER CT1940 [TML1711] and FIBER OPTIC CATHETER [HTF1491] form an Endovenous Laser Therapy (EVLT) solution designed for the treatment of varicose veins. The system delivers 1940nm laser energy through a fiber optic catheter directly to the target vein, enabling precise and effective vessel closure.
- With high tissue absorption efficiency, the system achieves optimal therapeutic outcomes using lower energy levels. The use of an ultra-thin catheter (≤ 0.8 mm) enables a minimally invasive procedure, reducing incision size, shortening recovery time, and supporting faster return to daily activities.
- EVLT system utilizing a 1940nm wavelength laser for endovenous treatment
- High tissue absorption enables efficient and precise vascular closure with reduced energy usage
- Advanced fiber optic technology ensures accurate laser delivery to the target vessel
- Ultra-thin 0.8 mm catheter minimizes incision size for a less invasive procedure
- Integrated temperature sensor automatically shuts off output in case of overheating
- Specialized cooling system prevents thermal overload even during high-power operation
- Standard SMA905 connector for reliable fiber optic coupling
- Supports reduced recovery time and faster patient return to daily life
Key Features
Compact Design & Efficient Cooling SystemA self-developed efficient heat dissipation system enables stable and effective laser performance within a compact device design.
Optical Fiber [Normal type]
Optical Fiber [Ball type]
Optical Fiber [Radial type]
Multiple Fiber Optic OptionsA wide range of fiber optic types is available to enhance procedural flexibility, allowing clinicians to select the optimal configuration based on patient vascular conditions and treatment objectives.
Product Usage Videos
Play Video
Technical Specifications
Laser wave length | 1940nm ± 30nm |
|---|---|
Laser Classification | Class IV (IEC60825-1:93 A2:2001) |
Laser type and output | Diodelaser, 0~10W(CW) |
Guide laser | 650nm, RED |
User interface | 7“ TFT LCD with touch panel |
Connector type | SMA905 |
User safety | Foot pedal, lock key, Emergency switch |
Overheat protection Temp. | over 35℃ |
Rated voltage | 100~240VAC |
Rated frequency | 50/60Hz |
PowerConsumption | 324VA |
Electric safety class | Class I, Type B |
Part No. | Length[mm] | Diameter |
|---|---|---|
HTF1491-6A | 1,800 | 0.8 |
HTF1491-6B | 2,200 | 0.8 |
HTF1491-6C | 2,800 | 0.8 |
HTF1491-4A | 2,800 | 0.6 |
HTF1491-B6A | 1,800 | 0.8 |
HTF1491-B6B | 2,200 | 0.8 |
HTF1491-B6C | 2,800 | 0.8 |
HTF1491-D6A | 2,200 | 0.7 |
HTF1491-D6B | 2,500 | 0.7 |
HTF1491-R6A | 2,200 | 0.7 |
HTF1491-R6B | 2,500 | 0.7 |